Box Medical
            Knowledge Base Box Medical Solutions Inc. Troubleshoot

            Exception Error: E0434352 - Software Freezing During Image Capture

            Update 06/13/2019:

            Software Freezing During Image Capture:

            1.  Exception Error:  E0434352
              •  This error can occur when a cable is loose in the insertion at the camera head.  After plugging in camera, secure cable into head by using a 1.5mm Hex Allen Screw to tighten the cable connection to the MX2 Camera.
              •  Check USB port and ensure that the port is a USB 3.0 slot
            2.  Confirm computer specifications and minimum RAM and Processing computer requirements are met.
            Updated: 27 Jun 2019 05:41 AM
            Helpful?  
            Help us to make this article better
            0 0