The hats should not be cleaned or reused, as they will stretch after single use. You do not need to clean the power cord. If the unit itself has significant debris on the outside, you can wipe it with a damp cloth.


    • Related Articles

    • How do I clean the accessories?

      The accessories can be cleaned by soaking them in a 0.5% bleach solution for 10 minutes (or follow your facility’s standard cleaning procedure). Bleaching time should not exceed one hour or the accessories may start to degrade. After bleaching, you ...
    • What Pumani accessories do I clean?

      The following items have been labeled as single-use only (nasal prongs, patient tubing, and fish elbows), and re-cleaning should be done at your own risk. The following items have been designed to be cleaned repeatedly and re-used: hat clips, tube ...
    • Can I use different accessories with the Pumani if I run out of accessories?

      No, the Pumani accessories are designed specifically to provide the correct levels of flow and pressure to the patient. If you use different accessories that are not part of the Pumani bubbleCPAP product, you may provide an unsafe level of flow or ...
    • Do I need a voltage converter with the Pumani?

      The Pumani runs on 220 volt AC (wall) power. If your facility uses 220 V AC power, you do not need a voltage converter. If your facility uses 120 V AC power, you do not need a voltage converter either. You may still need an adapter on the end of the ...
    • What do I need to do to make sure the Pumani device runs for many years?

      To optimize the performance of the Pumani unit, you may need to replace the pump filter every 1-2 years, depending on the level of debris that enters the unit. To check the functionality of the Pumani unit, determine if you need to replace the ...